dc.contributor.authorHah, Dooyoung
dc.contributor.authorJe, Chang Han
dc.contributor.authorLee, Sung-Q
dc.date.accessioned2021-07-28T10:36:33Z
dc.date.available2021-07-28T10:36:33Z
dc.date.issued2017en_US
dc.identifier.isbn978-1-5386-2952-9
dc.identifier.urihttps://hdl.handle.net/20.500.12573/879
dc.descriptionThis work was partially supported by Research Fund of the Abdullah Gul University (Project Number: FOA-2016-49).en_US
dc.description.abstractEffects of substrate bending to the characteristics of capacitive miniaturized ultrasonic transducers (CMUTs) on a flexible substrate are studied through finite element analysis (FEA) for the design purpose. The target application of the devices is intravascular ultrasonography (IVUS) where transducers are brought to the proximity of the imaging targets so that high resolution images can be obtained without much concern of signal attenuation. In order to eliminate mechanical rotation used in the conventional IVUS, the transducer array can be manufactured on a flexible substrate and to wrap it around a cylindrical frame. It can be anticipated that the characteristics of the transducers will be altered by such bending of the substrate through geometrical dimension changes and stress induced. Pullin voltages and resonant frequencies of CMUTs were studied via FEA for various bending radii and membrane thicknesses. It was found that both pull-in voltages and resonant frequencies become smaller for the transducers on a bent substrate compared to the ones on a flat substrate. It was also found that pull-in voltages decrease as the substrate bending radius is reduced.en_US
dc.description.sponsorshipAbdullah Gul University FOA-2016-49en_US
dc.language.isoengen_US
dc.publisherIEEE345 E 47TH ST, NEW YORK, NY 10017 USAen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectflexible substrateen_US
dc.subjectintravascular ultrasonography (IVUS)en_US
dc.subjectcapacitive miniaturized ultrasonic transducer (CMUT)en_US
dc.titleDesign of capacitive micromachined ultrasonic transducers (CMUTs) on a flexible substrate for intravascular ultrasonography (IVUS) applicationsen_US
dc.typeconferenceObjecten_US
dc.contributor.departmentAGÜ, Mühendislik Fakültesi, Elektrik - Elektronik Mühendisliği Bölümüen_US
dc.contributor.authorID0000-0002-1290-0597en_US
dc.relation.journal2017 SYMPOSIUM ON DESIGN, TEST, INTEGRATION AND PACKAGING OF MEMS/MOEMS (DTIP 2017)en_US
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanıen_US